Analyze Detect Measure Control
Users Guide
Redox/ORP
Electrodes
-
8/2/2019 electrodos mantenimiento
2/13
ROSS and the COIL trade dress are trademarks of Thermo Electron Corporation.
AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perpHect,
PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future,SensorLink, ROSS Ultra, Sure-Flow, TEA Analyzer, Titrator PLUS, TURBO2 and Wine
Master are registered trademarks of Thermo.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR,
AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR,
Auto-Test, BOD AutoE Z, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS,
digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra
Easy/Extra Value, FAST QC, Flash Titration, Flash Titrator, GAP, GLPcal, GLPcheck,GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir
Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch
Measurement, Optimum Results, Orion Constellation Software, Pentrode, pHuture
MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK,
rf link, ROSS, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART
STABILITY, Stacked, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH,
Unbreakable pH, Universal Access are trademarks of Thermo.
Guaranteed Success and The Technical Edge are service marks of Thermo.
PerpHecT meters are protected by U.S. patent 6,168,707.
PerpHecT ROSS are protected by U.S. patent 6,168,707.
ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093,
D334,208 and D346,753.
ionplus electrodes and Optimum Results solutions are protected byUS Patent 5,830,338.
ROSS Ultra electrodes are protected by US patents 6,793,787.
Orion ORP Standard is protected by US Patent 6,350,367.
Orion NoCal electrodes with stabilized potential patent pending.
Copyright 2005, Thermo Electron Corporation. All rights reserved. Questioneverything, and Analyze.Detect.Measure.Control are trademarks of Thermo
Electron Corporation.
The specifications, descriptions, drawings, ordering information and part numbers
within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
-
8/2/2019 electrodos mantenimiento
3/13
1Redox/ORP Electrode Manual
IntroductionThis manual contains instructions for the Orion Epoxy Sure-
Flow Combination Redox/ORP (9678BN), Glass Combination
Redox/ORP (9778BN), Epoxy Low Maintenance ORP Triode
(9179BN) and Epoxy Refillable ORP Triode (9180BN).
The Orion 9678BNWP and 9778BNWP combine a platinum
redox and a silver/silver chloride reference electrode in one
body. The Orion 9678BN is an epoxy body electrode with
a sleeve junction and the Orion 9778Bn has a glass body
with a ceramic junction. Both can be used to make direct
measurements of redox potentials or to perform potentiometric
titrations.
The Orion 9179BNMD and 9180BNMD consist of a platinum
redox sensor, a silver/silver chloride internal reference
system and a built-in thermistor for Automatic Temperature
Compensation (ATC). Orion 91-80 is a refillable electrode,
which uses Orion 900011 or 900001 filling solution. Orion 91-79
is a non-refillable, low maintenance gel-filled electrode. The
9179BNMD and 9180BNMD both have the capability to measureORP and Temperature simultaneously without a separate
reference electrode. Direct measurements of redox potentials or
potentiometric titrations can also be performed with these ORP
Triode electrodes.
Orion Orion
9179BNMD 9678BNWp
-
8/2/2019 electrodos mantenimiento
4/13
2 Redox/ORP Electrode Manual
All of the Orion Redox/ORP electrodes are not designed for
amperometric or constant current titrations, such as Karl
Fischer analysis.
It is strongly recommended to use ORP standards,
Cat. No. 967901 or 967961, to ensure accurate measurements.
Required Equipment1. Orion pH/mV meter (i.e. Orion 3 Star pH, Orion 4 Star pH/
ISE meter, Orion 5 Star multi-parameter)
Orion 290A, 290Aplus, 410A, 410Aplus, 420A, 420Aplus,
520A,520Aplus, 525A, 525Aplus, 710A, 710Aplus, 920A,
920Aplus will need to have an ATC adapter if you use
Triodes otherwise the other ORPs are interchanged
on these meter.
2. Orion Redox/ORP electrode
Orion Epoxy Sure-Flow Combination Redox/ORP
(9678BNWP), Glass Combination Redox/ORP
(9778BNWP), Epoxy Low Maintanence ORP
Triode (9179BNMD) and Epoxy Refillable ORP Triode
(9180BNMD)
3. Solutions
Electrode Electrode Calibration Standard ISAFill Solution
9678BNWP 900001 or 900011 967901 or 967961 None Required
9778BNWP 900001 or 900011 967901 or 967961 None Required
9179BNMD None Required 967901 or 967961 None Required
9180BNMD 900011 967901 or 967961 None Required
4. ATC temperature probe
5. Magnetic stirrer or stir bars are highly recommended
6. Beakers
7. Distilled or Deionized Water
-
8/2/2019 electrodos mantenimiento
5/13
3Redox/ORP Electrode Manual
Electrode Assembly and Preparation
Filling Instructions for Orion 9180BNMD,9778BNWOWP, 9678BNWP
1. Remove electrode protective cap and keep the cap for
storage.
2. Unplug the electrode fill hole cover on the electrode body.
3. Using the appropriate fill solution bottle, replace the flat
cap on the fill solution bottle with the flip-spout cap. Lift the
spout to the vertical position and insert into the fill hole.
4. Fill the electrode chamber up to a level just below the
fill hole.
5. Shake down the electrode as you would with a
thermometer to remove air bubbles.
6. After filling a dry chamber of Orion 9678BNWP (ONLY),
push cap and body together to leak some filling solution
past conical reference junction.
IMPORTANT:Do not place the electrode fill hole cover back
onto the electrode when making measurements!
Calibration Procedures
Manual ORP Calibration in Relative mV Mode:
The function of the relative mV mode is to calculate the mV
reading at a fixed user selected offset from absolute mV
measured from the Orion ORP electrode Orion 9179BNMD or
9180BNMD.
1. Set the meter to the relative millivolt mode. See Meter
Instruction Manual.
2. Place the ORP electrode in Orion ORP standard,
Orion 967901 or 967961.
Note:If using Orion 9180BNMD, make sure the proper filling
solution is used and filling instructions are followed.
-
8/2/2019 electrodos mantenimiento
6/13
4 Redox/ORP Electrode Manual
3. Calibrate the electrode by scrolling to the desired mV
value of the ORP solution. (mV values for EH are on the
bottle for Orion 967901 and see separate sheet for Orion
967961) See Meter Instruction Manual for more detailed
instructions.
4. Press YES key to accept value. All relative millivolt
measurements will be based on the offset value.
Note: If error message displays on main field, press MODE
key to abort calibration. See Troubleshooting Section in Meter
Instruction Manual.
ORP (EH) Autocalibration Procedure:
The function of the ORP mode is to calculate the equivalent
mV reading which would be read by a Standard Hydrogen
Electrode or Normal Hydrogen Electrode (NHE) at the measured
temperature using the measured input from the Orion ORP
electrode Orion 9179BNMD or 9180BNMD.
1. Press MODE key to set the meter to the EH mode. (First
time use of meter, dashes appear.)
2. Place the ORP electrode in Orion ORP standard, Orion
967901 or 967961.
Note: If using Orion 91-80, make sure the proper filling
solution is used and filling instructions are followed.
3. Press the 2nd key and then Cal key. When meter is ready,
the measured mV value which should be between the
range of +160 mV and +280 mV is displayed and converted
to the EH value.
Note: Error code E-21 will be displayed if mV is not
within this range. This error code is only a warning indicator.
Press MODE key to switch out of the ORP autocalibration
procedure to the mV mode. See Troubleshooting section.
4. Press YES key to accept the flashing EH value. Meter willautomatically switch back to the MEASURE mode.
-
8/2/2019 electrodos mantenimiento
7/13
5Redox/ORP Electrode Manual
Important Measurement Techniques Stir all standards and samples at a uniform rate during
measurement.
Magnetic stirrers may generate sufficient heat to change
solution temperature. Place a piece of insulating material
such as cork, cardboard or styrofoam between the stir plate
and sample beaker.
Always use fresh ORP standards for calibration.
Always rinse electrode(s) with distilled water thoroughly
between measurements. Gently shake electrode afterrinsing to prevent solution carry-over, then blot dry with a
lint-free tissue.
Allow all standards and samples to come to room
temperature for precise measurements.
Make sure the electrode filling solution is appropriate foryour sample. See Electrode Assembly and Preparation.
Electrode MeasurementsDirect Redox Potential (ORP) measurements can be used to
determine the oxidizing or reducing capability of a solution andare commonly made in polluted streams, chlorinated waters,
industrial waste waters, and other water samples.
1. After setting up the meter and calibrating the electrode,
the electrode is ready for measurements. When the meter
switches back to the MEASURE mode or is in relative mV
mode, place the electrode in the sample solution.
2. When the reading stabilizes, record the potential reading
and the temperature.
Note: Like all platinum redox electrodes, Orions 9179BNMD
and 9180BNMD may give unstable readings in solutions
containing chromous, vanadous, titanous, and other ions that
are stronger reducing agents than hydrogen or platinum.
Note:When requiring a specific measurement method
other than Direct Measurement, please refer to
www.thermo/water.com (i.e. Oxidation Reduction Titrations).
-
8/2/2019 electrodos mantenimiento
8/13
6 Redox/ORP Electrode Manual
Electrode Maintenance
Orion 9179BNMD
This Low-Maintenance ORP TriodeTM electrode can be rinsed
off with distilled water from any salt build-up.
Orion 9180BNMD
This refillable ORP Triode electrode should be routinely cleaned.
To remove precipitate from outside wall or on electrode tip,
rinse with distilled water. Make sure that the filling solution does
not crystallize or dry out. If crystallized, run the electrode under
distilled water or wipe the vent hole area with a damp tissueand then fill with filling solution. Always keep the electrode fill
hole plug on to prevent salt build up while storing for a short
period of time up to one day. Fill hole must be open during
measurements.
Electrode Cleaning Procedures
Orion 9678BNWP Cleaning
The electrode can be routinely cleaned without disassembling.
To remove precipitate that forms on the outside wall or tip of
the electrode, rinse with distilled water. If sample or precipitate
clogs the space between the electrode sleeve and the innercone, clean the chamber by flushing out the fill solution. To do
this, invert the electrode to moisten the O-ring. Holding the
electrode by the cap with one hand, push the outer sleeve up
into the cap with the other hand, allowing the fill solution to
drain from the chamber. If the chamber is not completely clean,
repeat refill with the fill solution.
If the preceding directions fail to remove all the precipitate, the
electrode can be disassembled and cleaned using the following
procedure: This procedure is not recommended for routine
cleaning of the electrode and should be used only when
absolutely necessary.
1. Unscrew the cap and slide the cap and epoxy coated springup along the cable. See Figure 1.
2. Holding the outer sleeve with one hand, firmly push down
on the threaded portion of the inner body to separate the
inner body from the sleeve.
-
8/2/2019 electrodos mantenimiento
9/13
7Redox/ORP Electrode Manual
3. Rinse the sleeve and the inner body thoroughly with
distilled water, and allow to air dry.
4. Moisten the O-ring on the electrode inner body with a drop
of fill solution. Reassemble and fill the electrode.
Changing the Fill SolutionThe electrode need not be disassembled to change the fill
solution in the reference chamber. Simply clean the electrode
and fill with the new solution, using the following procedure:
1. Tip the electrode to moisten the O-ring. Holding theelectrode by the cap with one hand, push the sleeve up into
the cap with the other hand allowing solution to drain from
the chamber.
2. Fill the electrode with distilled water and then drain the
reference chamber by retracting the sleeve, as explained
above. Now perform this filling and draining proceduretwice with the new filling solution.
3. Fill the electrode with new solution: Although the electrode
is ready for use, readings may drift slightly for about a half
hour while the electrode equilibrates with the new solution.
Orion 9778BNWP Cleaning
To remove precipate or salts that form on the exterior of the
electrode, rinse with distilled water. To remove salt deposits
formed inside the electrode:
1. Fill a wash bottle to half with distilled water.
2. Invert electrode so that glass membrane points up.
3. Invert the wash bottle so that its bottom points up.
4. Insert nozzle of wash bottle into electrode filling hole
and squeeze bottle. The fill solution will drain out ofthe chamber.
5. Fill chamber with distilled water from wash bottle.
-
8/2/2019 electrodos mantenimiento
10/13
8 Redox/ORP Electrode Manual
6. Continue to repeat steps 2-5 until all salts have been
removed.
To change the fill solution, follow steps 1-6 in cleaning above.
After draining out distilled water, fill with fill solution, drain, and
refill with the solution.
Orion 9179BN Cleaning
To remove precipitate or salts that form on the exterior of the
electrode, rinse with distilled water.
Orion 9180BN CleaningTo remove salt deposits formed inside the electrode, follow the
procedure below:
1. Fill a wash bottle to half with distilled water.
2. Invert electrode so that sensor points up.
3. Insert nozzle of wash bottle into electrode filling hole and
squeeze bottle.
4. Fill chamber with distilled water.
5. Take a 1 mL plastic disposable pipet and remove the liquidfrom the inverted electrode.
6. Repeat steps 2 to 4 until precipitate and deposits are
removed.
7. Rinse with fresh fill solution. Remove fill solution with pipet.
8. Fill with fresh fill solution.
-
8/2/2019 electrodos mantenimiento
11/13
9Redox/ORP Electrode Manual
Electrode Storage
Short-term Storage (up to one day):
Electrodes may be kept in distilled water or air. If left in air,
remove salt crystals on the outside of the electrode sleeve by
rinsing with distilled water.
Long-term Storage (more than 24 hours):
For Orion 9179BNMD, rinse electrode, dry with lint-free tissues
or air dry, and replace with electrode protective cap. Store DRY.
For Orion 9180BNMD, 9678BNWP, 9778BNWP, remove saltcrystals on the outside of the electrode by rinsing with distilled
water, drain filling solution from chamber, flush out with
distilled water and store dry. See Cleaning Procedures. Replace
electrode protective caps to prevent sensor damage.
TroubleshootingThe most important principle in troubleshooting is to isolate
the components of the system and check each in turn. The
components of the system are: 1) Meter 2) Electrode(s)
3) Standard and 4) Sample.
Meter
The meter is the easiest component to eliminate as a possible
cause of error. Orion meters are provided with an instrument
checkout procedure in the instruction manual and a shorting
cap for convenience in troubleshooting. Consult the manual for
complete instructions and verify that the instrument operates as
indicated and is stable in all steps.
Electrode(s)
Rinse electrode(s) thoroughly with distilled water.
Place electrode in the ORP solution and wait until the readingstabilizes.
The potential reading should be 220 60 mV. If the mV is out
of range, use fresh standards. If the problem persists, consult
with the Thermo technical service department.
-
8/2/2019 electrodos mantenimiento
12/13
10 Redox/ORP Electrode Manual
Standard
The quality of results depends greatly upon the quality of the
standards. ALWAYS prepare fresh standards when problems
arise it could save hours of frustrating troubleshooting!
SampleIf the electrodes work properly in standards but not in the
sample, look for possible interferences, or substances that could
affect response or physically damage the sensing electrode or
the reference electrode. If possible, determine the composition
of the samples and check for problems.
AssistanceAfter troubleshooting all components of your measurement
system, contact The Technical EdgeSM for Orion products. Within
the United States call 1.800.225.1480, outside the United States
call 978.232.6000 or fax 978.232.6031. In Europe, the Middle
East and Africa, contact your local authorized dealer. For the
most current contact information, visit www.thermo.com/water.
Warranty
For the most current warranty information,visit www.thermo.com/water.
-