Digital Weight Indicator
Installation Manual
8/12/2019 120 Manual de Servicio
2/41
8/12/2019 120 Manual de Servicio
3/41
2004 Rice Lak e Weighing Systems. All rights reserved. P rinted in the United States of America.
Specifications subject to change without notice.
February 2004
Technical training seminars are available through Rice Lake Weighing Systems.Course descriptions and dates can be viewed at www.rlws.comor obtained by calling
715-234-9171 and asking for the training department.
Contents
About This Manual ........................................................ ...................................................... ..................... 1
1.0 Introduction................................................................ ...................................................... ............ 1
1.1 O perating M odes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2 Front Panel K eypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.3 LED A nnunciators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.4 Indicator O perations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4.1 Weighing M ode O perations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4.2 Panel M ode O perations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.0 Insta lla tion............................................... ........................................................ ............................ 5
2.1 Unpacking and Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.2 Enclosure and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.1 Serial C ommunications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.2 Load Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Enclosure Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.4 R eplacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.0 Configuration................................................................................................... ............................. 7
3.1 C onfiguration M ethods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1.1 R evolution C onfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1.2 EDP C ommand C onfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1.3 Front Panel C onfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 M enu Structures and Parameter Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.1 C onfiguration M enu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.2 Format M enu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2.3 C alibration M enu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.4 Serial M enu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2.5 Program M enu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2.6 Print Format M enu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2.7 Time M enu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.8 Date M enu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.9 Version M enu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.0 Calibration................................................ ........................................................ .......................... 21
4.1 Front Panel C alibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214.2 EDP C ommand Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.0 EDP Commands................................................. ........................................................ ................. 23
5.1 T he EDP C ommand Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.1 K ey Press C ommands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.2 M ode C ommands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.3 R eporting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.4 The RS C ommand. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.5 Parameter Setting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2 Saving and Transferring Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.2.1 Saving Indicator Data to a Personal C omputer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.2.2 Downloading Configuration Data from PC to Indicator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8/12/2019 120 Manual de Servicio
4/41
ii
120 Installation Manual
6.0 Print Formatting ............................................... ........................................................ .................. 27
6.1 P rint Formatting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 276.2 C ustomizing Print Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.2.1 Using the EDP P ort. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.2.2 Using the Front Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.0 Appendix .................................................. ........................................................ .......................... 29
7.1 Error M essages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.2 C ontinuous O utput (Stream) Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 297.3 Front Panel Display C haracters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 307.4 A SC II Character C hart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 317.5 C onversion Factors for Secondary Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337.6 D igital Filtering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.6.1 DIG FL
x
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.6.2 DFSENS and DFTH RH Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.6.3 Setting the Digital Filter Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.7 T est M ode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357.8 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
120 Limited Warranty....................................................... .................................................... .................. 37
8/12/2019 120 Manual de Servicio
5/41
Introduction
1
About This Manual
This manual is intended for use by service techniciansresponsible for installing and servicing 120
digitalweight indicators. This manual applies to indicatorsusing Version 1.0 of the 120
software.
Conguration and calibration of the indicator can beaccomplished using the indicator front panel keys, theEDP command set, or Version 3.0 or later of theRevolution
conguration utility. See Section 3.1 onpage 7for information about conguration methods.
Some procedures described in this
manual require work inside the
indicator enclosure. These proc edures
are to be perfo rmed by qualified
service personnel only.
Authorized distributors and their employeescan view or download this manual from theRice Lake Weighing Systems distributorsite at www.rlws.com
.
The Operator Card
included with this manualprovides basic operating instructions for users of the
120
. Please leave the Operator Card
with the indicatorwhen installation and conguration are complete.
1.0 Introduction
The 120
is a single-channel digital weight indicator housed in a durable plastic enclosure. The indicator frontpanel consists of a large (.8 in, 20 mm), six-digit, seven-segment LED display and ve-button keypad. Featuresinclude:
Drives up to four 350
W or eight 700
W
load cells Supports 4- and 6-wire load cell connections Electronic data processing (EDP) port for full duplex, RS-232 communications at up to 38400 bps Printer port for output-only RS-232 and 20 mA current loop communications at up to 9600 bps
The 120
is NTEP-certied for Classes III and III L at 6,000 divisions. See Section 7.8 on page 36for detailedspecications.
1.1 Operating Modes
The 120
supports the following modes of operation:
Normal (weighing) mode
Normal mode is the production mode of the indicator. The indicator displays gross or net weights asrequired, using the LED annunciators described in Section 1.3 on page 2to indicate scale status and the typeof weight value displayed. Once conguration is complete and a legal seal is afxed to the back of theindicator, this is the only mode in which the 120
can operate. See Section 1.4.1 on page 3 for moreinformation about normal mode operations.
Pa nel mode
Panel mode allows the time, date, consecutive number, and consecutive number start-up value to be setwithout entering conguration mode. To enter panel mode, press and hold the GROSS/NET
key until the
TIME
menu is shown. See Section 1.4.2 on page 4for more information about panel mode.
Configuration mode
Most of the procedures described in this manual require the indicator to be in conguration mode, includingconguration and calibration.To enter conguration mode, remove the large llister head screw from the enclosure backplate. Insert ascrewdriver or a similar tool into the access hole and press the setup switch once. The indicator displaychanges to show the word CONFIG
.
Test mode
Test mode provides a number of diagnostic functions for the 120
indicator. Like setup mode, test mode isentered using the setup switch. See Section 7.7 on page 35for more information about entering and usingtest mode.
Warning
8/12/2019 120 Manual de Servicio
6/41
2
120 Installation Manual
1.2 Front Panel Keypad
Figure 1-1shows the 120
keypad and LED annunciators.
The symbols shown under the keys (representing up, down, enter, left, right) describe the key functions assignedin conguration and panel modes. In these modes, the keys are used to navigate through menus, select digitswithin numeric values, and increment/decrement values. See Section 3.1.3 on page 8for information about usingthe front panel keys in conguration mode.
Figure 1-1. 120
Front Panel
1.3 LED Annunciators
The 120
display uses a set of eight LED annunciators to provide additional information about the value beingdisplayed:
Gr
(gross) and Nt
(net) annunciators are lit to show whether the displayed weight is a gross or net weight. Center of zero ( ): Gross weight is within 0.25 graduations of zero. This annunciator lights when the
scale is zeroed.
Standstill ( ): Scale is at standstill or within the specied motion band. Some operations, includingtare functions and printing, can only be done when the standstill symbol is shown.
lb, kg, oz, and
g annunciators indicate the units associated with the displayed value: lb=pounds,kg=kilograms, oz=ounces, g=grams.The displayed units can also be set to short tons (tn), metric tons (t), or NONE (no units informationdisplayed). The lb
and kg
LEDs function as primary and secondary units annunciators for somecombinations of primary and secondary units. If neither primary nor secondary units are lb, kg, oz, or g, the
lb
annunciator is lit for primary units, kg
for secondary units.Table 1-1 on page 3shows which annunciators are used for all combinations of congured primary andsecondary units. For example:
If the primary unit is pounds (lb) and the secondary unit is kilograms (kg), the lb
LED is lit for primaryunits, kg
for secondary units. If the primary unit is pounds (lb) and the secondary unit is short tons (tn), the lb
LED is lit for primaryunits, kg
for secondary units. There is no LED for short tons, so the kg
LED is used as the secondaryunits annunciator.
If the primary unit is short tons (tn) and the secondary unit is pounds (lb), the lb
LED is lit for primaryunits (tn), and kg
is lit for secondary units (lb). Because there is no LED for short tons, the lb
and kg
LEDs are used as primary and secondary units annunciators.
kgl b goz Gr Nt
ZERO
GROSS
NET
B/N
TARE
T
UNITS
Units
8/12/2019 120 Manual de Servicio
7/41
Introduction
3
See Section 3.2.2 for more information about conguring primary and secondary display units.
1.4 Indicator Operations
1.4.1 Weighing Mode Operations
Basic 120
operations are summarized below:
Toggle Gross/Net Mode
Press the GROSS/NET
key to switch the display mode from gross to net, or from net to gross. If a tare value hasbeen entered or acquired, the net value is the gross weight minus the tare.
Gross mode is shown by the Gross
annunciator; net mode is shown by the Net
annunciator.
Toggle Units
Press the UNITS
key to switch between primary and secondary units. The units LED to the right of the display islit.
Zero Sca le
1. In gross mode, remove all weight from the scale and wait for the standstill annunciator ( ).2. Press the ZERO
key. The center of zero ( ) annunciator lights to indicate the scale is zeroed.
Acquire Tare
1. Place container on scale and wait for the standstill annunciator ( ).2. Press the TARE
key to acquire the tare weight of the container. The indicator switches to net mode.
Remove S tored Tare Value
1. Remove all weight from the scale and wait for the standstill annunciator ( ).2. Press the TARE
key. The indicator switches to gross mode, indicating the tare value has been removed.
Print Ticket
1. Wait for the standstill annunciator ( ).2. Press the PRINT
key to send data to the serial port.
Primary Unit
Secondary Unit
lb kg oz g tn t none
lb lb / lb lb / kg lb / oz lb / g lb / kg
kg kg / lb kg / kg kg / oz kg / g lb / kg
oz oz / lb oz / kg oz / oz oz / g oz / kg
g g / lb g / kg g / oz g / g g / kg
tn lb / kg lb / kg lb / oz lb / g lb / lb lb / kg lb / kg
t lb / kg lb / lb lb / kg
none lb / kg lb / kg lb / lb
Table 1-1. Units Annunciators, Showing Primary / Secondary LEDs Used for All Congurations
8/12/2019 120 Manual de Servicio
8/41
4
120 Installation Manual
1.4.2 Panel Mode Operations
The following operations are available by placing the indicator in panel mode:
Set time Set date Set consecutive number
Set consecutive number start-up valueTo enter panel mode, press and hold the GROSS/NET
key until the TIME
menu is displayed. Use the navigationkeys to move around the menu; to change a value, use the navigation keys to select the digit and increment ordecrement its value. Press the Enter
(
TARE
) key to set the value and return to the menu level above. Figure 1-2shows the structure of the panel mode menu.
Figure 1-2. Panel Mode Menu Structure
T IME
SHOW
hh.mm.ss
HOUR
hh
MINUTE
mm
SECOND
ss
DATE
SHOW
yy.mm.dd
YEAR
yy
M O N T H
mm
DAY
dd
CONSNU
number
CONSTU
number
8/12/2019 120 Manual de Servicio
9/41
Installation
5
2.0 Installa tion
This section provides information for connecting load cell and serial communications cables to the 120indicator.
2.1 Unpacking and Assembly
Immediately after unpacking, visually inspect the 120 to ensure all components are included and undamaged.The shipping carton should contain the indicator with attached tilt stand, this manual, and a parts kit. If any partswere damaged in shipment, notify Rice Lake Weighing Systems and the shipper immediately.
The parts kit contains the items listed below:
Capacity and identication labels Load cell connector (PN 82505) 9V power supply adapter (PN 78611 for 115 VAC units, PN 78612 for 230 VAC units)
2.2 Enclosure a nd ConnectorsThe back of the 120enclosure provides a 3-pin power connection, 9-pin D-sub connector for communications,and a 6-pin connector for load cell connection (see Figure 2-1).
The setup switch, used for placing the indicator into conguration mode, is located in the recess on the undersideof the enclosure. The setup switch is protected by a cover plate and secured with a llister head screw (not shownin Figure 2-1).
Figure 2-1. Back View of 120Enclosure, Showing Load Cell and Communications Connectors and Setup Switch Location
SETUPPORT 1 PORT 2
COMMUNICATIONS LOAD CELL
9VDC ADAPTER
SWITCH
8/12/2019 120 Manual de Servicio
10/41
6 120 Installation Manual
2.2.1 Serial Communications
The serial communications cable attaches to the maleD-Sub connector, Port 1 (see Figure 2-1 on page 5).Port 1 provides connections for the EDP (ElectronicData Processing) port and the printer port. Table 2-1.shows the pin assignments for Port 1.
The EDP port supports RS-232 communications only;the printer port provides either active 20 mA output orRS-232 transmission. Both ports are congured usingthe SERIAL menu. See Section 3.0 on page 7forconguration information.
2.2.2 Loa d Cells
The load cell or junction box cable attaches to theround 6-pin connector, Port 2 (see Figure 2-1 on page5). Table 2-2shows the pin assignments for Port 2.If using 4-wire load cell cable (without sense wires),short pin 2 to pin 3, and pin 4 to pin 5.
2.3 Enclosure Disassembly
Use a wrist strap to ground yourself and protect
components from electrostatic discharge (ESD) when
working inside the indicator enclosure.
If the indicator enclosure must be opened formaintenance, do the following:
1. Disconnect power to the unit. Remove tilt stand.2. Remove two llister head screws and the setup
switch cover plate from back of enclosure.3. Loosen self-tapping screw at top center of back
of enclosure.4. Lift up the forward edge of the rubber feet on
bottom of enclosure for access to two additionalself-tapping screws. Loosen both screws.
5. Press down on top of back half of the enclosureto release tabs. Open enclosure by separating thehousing at the top of the indicator. (CPU boardis mounted to front half of enclosure; power,communications, and load cell connections allconnect to the bottom of the CPU board.)
6. Reverse steps to reassemble enclosure.
2.4 Replacement Parts
Table 2-3lists replacement parts for the 120indicator.
Port 1
P in Port Function
1 Printer RS-232 TxD
2 EDP RS-232 TxD
3 RS-232 RxD
4 not used
5 ED P/P rinter R S-232 G round / 20 mA O U T
6 N/C not used
7
8
9 Printer +20 mA O UT
Table 2-1. Serial Connector (Port 1) Pin Assignments
Port 2 P in Function
1 +SIG
2 +EXC
3 +SENSE *
4 EXC
5 SENSE *
6 SIG
* For 4-wire connections, short pin 2 to pin 3, pin 4 to pin 5.
Table 2-2. Load Cell Connector (Port 2) Pin Assignments
CautionPN Description
78609 C PU Board
78610 Switch panel membrane
15799 9-pin socket for D-sub communications cable
15774 Shell for D-sub communications cable
83429 Setup switch cover plate
83430 Fillister head screw
83432 Self-tapping screw (enclosure)
83431 R ubber foot
83428 Tilt stand wing knob
78949 O ptional wall-mount tilt stand
78611 9V power supply adapter for 115V units
78612 9V power supply adapter for 230V units
Table 2-3. Replacement Parts
8/12/2019 120 Manual de Servicio
11/41
Configuration 7
3.0 Configura tion
To congure the 120indicator, the indicator must be placed in setupmode. The setup switch is accessed by removing the left llister headscrew on underside of the enclosure and removing the rectangularswitch cover plate. Switch position is changed by inserting a
screwdriver into the access hole and pressing the switch.When the indicator is placed in setup mode, the word CONFIGisshown on the display. The CONFIG menu is the rst of nine mainmenus used to congure the indicator. Detailed descriptions of thesemenus are given in Section 3.2. When conguration is complete,return to the CONFIG menu and press the (ZERO) key to exit setup mode, then replace the setup switch accessscrew.
3.1 Configuration MethodsThe 120indicator can be congured by using the front panel keys to navigate through a series of congurationmenus or by sending commands or conguration data to the EDP port. Conguration using the menus isdescribed in Section 3.1.3. Conguration using the EDP port can be accomplished using the EDP command set
described in Section 5.0or by using the Revolution IIIconguration utility.3.1.1 Revolution Configuration
The Revolution IIIconguration utility provides thepreferred method for conguring the 120 indicator.Revolution runs on a personal computer to setconguration parameters for the indicator. WhenRevolution conguration is complete, congurationdata is downloaded to the indicator.
Figure 3-1. Sample Revolution Conguration Display
R e v o l u t i o n I I I supports both uploading anddownloading of indicator conguration data. Thiscapability allows conguration data to be retrievedfrom one indicator, edited, then downloaded toanother.
To use Revolution III, do the following:
1. Install Revolution III(Version 3.1 or later) on anIBM-compatible personal computer runningWindows 98 or later. Minimum systemrequirements include a processor speed of atleast 166MHz, 32MB of memory (64MBrecommended, required for NT4, 2000, XP),and at least 40MB of available hard disk spacefor installation.
2. With both indicator and PC powered off,connect the PC serial port to the RS-232 pins on
the indicator EDP port.3. Power up the PC and the indicator. Use the setup
switch to place the indicator in setup mode.4. Start the Revolution IIIprogram.
Figure 3-1 shows an example of one of the Revolutionconguration displays.
Revolution IIIprovides online help for each of i tsconguration displays. Parameter descriptionsprovided in this manual for front panel congurationcan also be used when conguring the indicator usingRevolution: the interface is different, but theparameters set are the same.
SETUP SWITCH
FILLISTER-HEAD SCR EW
8/12/2019 120 Manual de Servicio
12/41
8 120 Installation Manual
3.1.2 EDP Command Configuration
The EDP command set can be used to congure the 120indicator using a personal computer, terminal, or remotekeyboard. Like Revolution, EDP command conguration sends commands to the indicator EDP port; unlikeRevolution, EDP commands can be sent using any external device capable of sending ASCII characters over aserial connection.
EDP commands duplicate the functions available using the indicator front panel and provide some functions not
otherwise available. EDP commands can be used to simulate pressing front panel keys, to congure the indicator,or to dump lists of parameter settings. See Section 5.0 on page 23for more information about using the EDPcommand set.
3.1.3 Front Panel Configuration
The 120indicator can be congured using a series of menus accessed through the indicator front panel when theindicator is in setup mode. Table 3-1summarizes the functions of each of the main menus.
Four front panel keys are used as directional keys to navigate through the menus in setup mode. The UNITS( )and PRINT ( ) keys scroll left and right (horizontally) on the same menu level; ZERO( ) and GROSS/NET( )move up and down (vertically) to different menu levels. The TAREkey ( ) serves as an Enter key for selectingparameter values within the menus. A label under each of these keys identies the direction provided by the keywhen navigating through the setup menus.
To select a parameter, press or to scroll left or right until the desired menu group appears on the display,then press to move down to the submenu or parameter you want. When moving through the menu parameters,the default or previously selected value appears rst on the display.
To change a parameter value, scroll left or right to view the values for that parameter. When the desired valueappears on the display, press to select the value and move back up one level.
NOTE:You must press to save the selected value. The 120d oes not automatically save the last-displayed value.
To edit numerical values, press (rightmost digit willash), then use the navigation keys to select the digit
and to increment or decrement the value (seeFigure 3.2). When done, press again to save theedited value.
Figure 3-2. Editing Procedure for Numeric Values
M enu M enu Function
C O NFIG C onfiguration C onfigure load cell sensitivity, grads, zero tracking, zero range, motion band, overload, sample
rate, and digital filtering parameters.
FOR M AT Format Set format of primary and secondary units, display rate.
C ALIB R C alibration C alibrate indicator. S ee Section 4.0 on page 21for calibration procedures.
SER IAL Serial C onfigure EDP and printer serial ports.
PR O G RM Program Set power-up mode, regulatory mode, and consecutive number values.
P FO RM T Print Format Set print format used for gross and net tickets. See Section 5.0for more information.
T IM E T ime Display and set time
DATE Date Display and set date
VERS ION Version D isplay installed software version number.
Table 3-1. 120Menu Summary
When editing numeric values, press or to change the
digit selected. P ress or to increment or decrement the
value of the selected digit.
Press to save the value entered and return to the level above.
000 0 0 0
8/12/2019 120 Manual de Servicio
13/41
Configuration 9
3.2 Menu Structures a nd Parameter DescriptionsThe following sections provide graphic representations of the 120menu structures. In the actual menu structure,the settings you choose under each parameter are arranged horizontally. To save page space, menu choices areshown in vertical columns. The factory default setting appears at the top of each column.
Most menu diagrams are accompanied by a table that describes all parameters and parameter values associatedwith that menu. Default parameter values are shown in bold type.
3.2.1 Configuration Menu
Figure 3-3. Conguration Menu
GRADS
10000 OFF
1D
0.5D
3D
ZTRK BN ZRANG E
1.9%
100%
50D
2D
1D
3D
M O TB A N
10D
5D
20D
FS+2%
FS+9D
FS+1D
FS
OVRLOA
8
4
DIGFL1
64
32
number
DIG FL2 DIG FL3
8OUT
32O UT
16O UT
64O UT
DFSENS DFTHRH
1OUT
128OU T
2OUT
2DD
NONE
5DD
20DD
10DD
50DD
200DD
100DD
250DD
30HZ
7.5HZ
15HZ
3.75HZ
SMPRAT
PG A
2
4
O FF
1
2
4OUT
16
8
4
64
32
1
2
16
128 128
VERSDATET IMEPFORMTSERIALCALIBRC O NFIG FO RM AT P R O G R M
8
4
64
32
1
2
16
128
8
1
0.1DD
0.2DD
1DD
0.5DD
8/12/2019 120 Manual de Servicio
14/41
10 120 Installation Manual
CONFIG Menu
Parameter Choices Description
Level 2 submenus
PG A 2
4
8
1
Set programmable gain amplifier for load cell sensitivity:
0.5 mV/V to 4.5 mV/V PG A = 1
0.2 mV/V to 2.2 mV/V PG A = 2
0.1 mV/V to 1.1 mV/V PG A = 3
0.5 mV/V to 0.6 mV/V PG A = 4
GRADS 10000
number
G raduations. Specifies the number of full scale graduations. The value entered must be in
the range 1100 000 and should be consistent with legal requirements and environmental
limits on system resolution.
To calculate G R AD S, use the formula, G RAD S = Capacity/ Display Divisions.
D isplay divisions for primary and secondary units are specified on the FO R M AT menu.
ZTRKBN OFF
0.5D
1D
3D
Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the configured zero range (ZR ANG E parameter). Selections are
display divisions. M aximum legal value varies depending on local regulations.
ZRANGE 1.9%
100%
Zero range. Selects the range within which the scale can be zeroed. T he 1.9% selection is
1.9% around the calibrated zero point, for a total range of 3.8% . Indicator must be at
standstill to zero the scale. U se 1.9% for legal-for-trade applications.
M O T B A N 1D
2D
3D
5D
10D
20D
50D
O FF
M otion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected for 1 second or more, the standstill symbol lights. Some
operations, including print, tare, and zero, require the scale to be at standstill. M aximum
legal value varies depending on local regulations.
If O FF is selected, ZTRK BN should also be set to OFF.
OVRLOA FS+2%FS+1D
FS+9D
FS
O verload. Determines the point at which the display blanks and an out-of-range errormessage is displayed. M aximum legal value varies depending on local regulations.
SMPRAT 15HZ
7.5HZ
3.75HZ
30HZ
Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital
converter. Lower sample rate values provide greater signal noise immunity.
DIGFL1
DIGFL2
DIGFL3
2
4
8
16
32
64
1
Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanical
vibration from the immediate area of the scale.
C hoices indicate the number of A/D conversions that are averaged to obtain the
displayed reading. A higher number gives a more accurate display by minimizing the
effect of a few noisy readings, but slows down the settling rate of the indicator. See
Section 7.6 on page 34for more information on digital filtering.
DFSENS 8OUT
16O UT
32O UT
64O UT
128OUT
2OUT
4OUT
Digital filter cutout sensitivity. S pecifies the number of consecutive readings that must fall
outside the filter threshold (DFTHRH parameter) before digital filtering is suspended. If
NO NE is selected, the filter is always enabled.
Table 3-2. Conguration Menu Parameters
8/12/2019 120 Manual de Servicio
15/41
Configuration 11
DFTHRH NONE
0.1DD
0.2DD
0.5DD
1DD
2DD
5DD
10DD
20DD
50DD
100DD
200DD
250DD
Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When a
specified number of consecutive scale readings (DFSEN S parameter) fall outside of this
threshold, digital filtering is suspended. If N O NE is selected, the filter is always enabled.
CONFIG Menu
Parameter Choices Description
Table 3-2. Conguration Menu Parameters (Continued)
8/12/2019 120 Manual de Servicio
16/41
12 120 Installation Manual
3.2.2 Format Menu
Figure 3-4. Format Menu
FORMATMenu
Parameter Choices Description
Level 2 submenus
PR IM AR DEC PNT
DSPDIV
UNITS
Specifies the decimal position, display divisions, and units used for the primary units. See
Level 3 submenu parameter descriptions.
SEC NDR DEC PNT
DSPDIV
UNITS
M ULT
M ULEXP
Specifies the decimal position, display divisions, units, and conversion multiplier used for
the secondary units. See Level 3 submenu parameter descriptions.
DSPRAT 250MS
500M S
750M S
1SEC
1500M S
2SEC
2500M S
3SEC4SEC
6SEC
8SEC
Display rate. Sets the update rate for displayed values. Values are in milliseconds (M S) or
seconds (SEC ).
Table 3-3. Format Menu Parameters
DSPDIV M ULTUNITSDECPNT
0.45359
1D
5D
2D
number
PR IM AR SEC NDR
6SEC
4SEC
3SEC
2500M S
2SEC
1500M S
750M S
500M S
250MS
8SEC
DSPRAT
DSPDIV UNITSDECPNT
T
K G
1D
5D
2D
LB
8.88888
888888
88888.8
8888.88
888.888
88.8888
O Z
TN
NONE
G
KG
LB
M ULEXP
1SEC8.88888
888888
88888.8
8888.88
888.888
88.8888
dec_position
VERSDATET IMEPFOR MTSERIALCALIBRC O NFIG FO R M AT P R O G R M
20D
10D
50D
20D
10D
50D T
O Z
TN
NONE
G
8/12/2019 120 Manual de Servicio
17/41
Configuration 13
Level 3 submenus
Primary Units (PRIMAR Parameter)
DECPNT 888888
8.88888
88.8888
888.888
8888.88
88888.8
Decimal point location. Specifies the location of the decimal point or dummy zeroes in the
primary unit display. Value should be consistent with local legal requirements.
DSPDIV 1D
2D
5D
10D
20D
50D
Display divisions. Selects the minimum division size for the primary units displayed weight.
UNITS LB
K G
O ZTN
T
G
N O N E
Specifies primary units for displayed and printed weight. Values are: LB = pound;
K G = kilogram; O Z= ounce; TN = short ton; T = metric ton; G = gram.
Secondary Units (SECNDR Parameter)
DECPNT 88888.8
888888
8.88888
88.8888
888.888
8888.88
Decimal point location. D etermines the location of the decimal point or dummy zeros in
the secondary unit display.
DSPDIV 5D
10D
20D
50D
1D
2D
Display divisions. Selects the value of minimum division size of the displayed weight.
UNITS KG
O Z
TN
T
G
LB
N O N E
Specifies secondary units for displayed and printed weight. Values are: K G = kilogram;
O Z= ounce; TN = short ton; T = metric ton; G = gram; LB = pound.
M ULT 0.45359
number
M ultiplier. S pecifies the conversion factor by which the primary units are multiplied to
obtain the secondary units. T he default is 0.45359, which is the conversion factor forchanging pounds to kilograms. Use the M ULEXP parameter to shift the decimal position
of the multiplier. See Section 7.5 on page 33 for a list of multipliers.
To toggle between primary and secondary units, press the UN ITS key.
M ULEXP dec_position M ultiplier decimal shift. S pecifies a divisor used to shift the decimal position in the
secondary units multiplier value. U se the left and right arrow keys to shift the decimal
point within the displayed M ULT value.
FORMATMenu
Parameter Choices Description
Table 3-3. Format Menu Parameters (Continued)
8/12/2019 120 Manual de Servicio
18/41
14 120 Installation Manual
3.2.3 Calibration Menu
See Section 4.0 on page 21for calibration procedures.
Figure 3-5. Calibration Menu
CALIBR Menu
Parameter Choices Description
Level 2 subm enus
WZER O D isplay and edit the zero calibration A/D count value.
DO NOT ad just this value after WSPAN has been set!
WVAL 10000
test_weight
Display and edit the test weight value.
WSPAN Display and edit the span calibration A/D count value.
REZERO Press Enter to remove an offset value from the zero and span calibrations.
Use this parameter only after WZER O and WSPAN have been set. See Section 4.1 on
page 21 for more information about using this parameter.
Table 3-4. Calibration Menu Parameters
WZERO
*CAL*Display and editzero calibrationA/D count value
Display and edittest weight value
WVAL
*CAL*Display and editspan calibrationA/D count value
WSPAN
*CAL*Press Enter to
remove offset fromzero and spancalibrations
REZERO
VERSDATET IMEPFORM TSERIALCALIBRC O NFIG FO R M AT P R O G R M
10000
8/12/2019 120 Manual de Servicio
19/41
Configuration 15
3.2.4 Seria l Menu
See Section 7.2 on page 29for information about the 120serial data format.
Figure 3-6. Serial Menu
EDP PR INT
P R N
EDP
PRNDES
BITS TER MBAUD
C R
9600
7SPACE
7ODD
8NONE CR/LF
1200
38400
19200
EDP
OFF
STREAM
P RN
VERSDATET IMEP F OR M TSERIALCALIBRC O NFIG FO R M AT
DISABLE
ENABLE
P R OT C T
2SEC
1SEC
STM DLY
4SEC
4800
2400
7EVEN
BITS TER MBAUD
C R
9600
7SPACE
7ODD
8NONE CR/LF
1200
4800
2400
7EVEN
P R O G R M
15SEC
8SEC
N ON E
250MS
500M S
ECHO
O FF
ON
SERIAL Menu
Parameter Choices Description
Level 2 subm enus
EDP BAUD
BITSTERM
Specifies settings for baud rate, data bits, termination characters, and end-of-line delay used by
the EDP port.
PR INT BAUD
BITS
TERM
Specifies settings for baud rate, data bits, termination characters, and end-of-line delay used by
the printer port.
STREAM OFF
EDP
PRN
Selects the serial port used for continuous transmission. See Section 7.2 on page 29 for
information about the 120continuous data format.
Table 3-5. Serial Menu Parameters
8/12/2019 120 Manual de Servicio
20/41
16 120 Installation Manual
STM DLY 250MS
500M S
1SEC
2SEC
4SEC
8SEC
15SEC
NONE
Stream delay. Specifies the delay, seconds (SEC ) or milliseconds (M S), inserted between stream
frames.
PRNDES EDP
PRN
Print destination. Selects the port for data transmission when the PR IN T key is pressed or the
K PR INT EDP command is sent.
P R O T C T ENABLE
DISABL
EDP port protection. Select ENABLE to secure the EDP port.
Level 3 Submenus EDP Port
BAUD 9600
19200
38400
12002400
4800
Baud rate. Selects the transmission speed for the EDP port.
BITS 8NONE
7ODD
7EVEN
7SPACE
Selects number of data bits and parity of data transmitted from the EDP port.
TERM CR/LF
C R
Termination character. Selects termination character for data sent from the ED P port.
ECHO ON
O FF
Echo. Specify whether serial commands sent to the indicator are echoed.
Level 3 Submenus Printer Port
BAUD 96001200
2400
4800
Baud rate. Selects the transmission speed for the printer port.
BITS 8NONE
7ODD
7EVEN
7SPACE
Selects number of data bits and parity of data transmitted from the printer port.
TERM CR/LF
C R
Termination character. Selects termination character for data sent from the printer port.
SERIAL Menu
Parameter Choices Description
Table 3-5. Serial Menu Parameters (Continued)
8/12/2019 120 Manual de Servicio
21/41
Configuration 17
3.2.5 Program Menu
Figure 3-7. ProgramMenu
PWRUPM
GO NTEP
O I M L
REGULA
0
number
CONSNU
DELAY
0
number
CONSTU
CANADA
N ON E
VERSDATET IMEP F OR M TSERIALCALIBRC O NFIG FO R M AT P R O G R M
PROGRM Menu
Parameter Choices Description
Level 2 submenus
PWRUPM GO
DELAY
Power up mode. In G O mode, the indicator goes into operation immediately after a brief power
up display test.
In DELAY mode, the indicator performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset and
the warm up period repeated.
REGULA NTEP
O I M L
CANADA
N O N E
R egulatory mode. S pecifies the regulatory agency having jurisdiction over the scale site.
O IM L, NT EP, and CAN AD A modes allow a tare to be acquired at any weight greater than
zero. NO NE allows tares to be acquired at any weight value.
O IM L, NTEP, and C AN ADA modes allow a tare to be cleared only if the gross weight is atno load. N O NE allows tares to be cleared at any weight value.
NTEP and O IM L modes allow a new tare to be acquired even if a tare is already present.
In C ANADA mode, the previous tare must be cleared before a new tare can be acquired.
NO NE, NTEP and CA NAD A modes allow the scale to be zeroed in either gross or net
mode as long as the current weight is within the specified ZR AN G E. In O IM L mode, the
scale must be in gross mode before it can be zeroed; pressing the ZER O key in net mode
clears the tare.
CONSNU 0
number
C onsecutive numbering. A llows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.
The initial value of this parameter is set to the start up value specified on the CO NSTU
parameter. C hanging either C O NST U or CO NSN U immediately resets the consecutive number
used for printing.
CONSTU 0number
C onsecutive number start up value. Specifies the initial consecutive number (CO NSNU) valueused when the indicator is powered on.
Table 3-6. Program Menu Parameters
8/12/2019 120 Manual de Servicio
22/41
8/12/2019 120 Manual de Servicio
23/41
Configuration 19
3.2.7 Time Menu
Time can also be set by the operator in panel mode. See Section 1.4.2 on page 4.
Figure 3-9. Time Menu
3.2.8 Da te Menu
Date can also be set by the operator in panel mode. See Section 1.4.2 on page 4.
Figure 3-10. Date Menu
TIME Menu
Parameter Choices Description
Level 2 submenusSHOW HH.MM.SS Displays current time in HH.M M .SS format
H O U R hour (HH) Set hour using 24-hour format.
M INUTE minute (MM ) Set minute
SECOND second (SS) Set second
Table 3-7. Time Menu Parameters
DATE Menu
Parameter Choices Description
Level 2 submenus
SHOW YY.MM.DD .D isplays current date in YY. M M .D D format
YEAR year (YY) Set year (two digits, 0099)
M O N T H month (MM) Set month
DAY day (DD) Set day
Table 3-8. Date Menu Parameters
SHOW
00.00.00
HO UR M INUTE
display time 0059
SECOND
VERSDATET IMEP F OR M TSERIALCALIBRC O NFIG FO R M AT P R O G R M
00590023
00 00 00
SHOW
03.01.01
Y EAR M O NTH
display date 01-31
DAY
VERSDATET IMEP F OR M TSERIALCALIBRC O NFIG FO R M AT P R O G R M
01120099
03 01 01
8/12/2019 120 Manual de Servicio
24/41
20 120 Installation Manual
3.2.9 Version Menu
The VERS menu is used to check the software version installed in the indicator. There are no parametersassociated with the Version menu: when selected, the indicator displays the installed software version number.and the indicator model
Figure 3-11. Version Menu
Softwareversion
VERSDATET IMEP F OR M TSERIALCALIBRC O NFIG FO R M AT P R O G R M
8/12/2019 120 Manual de Servicio
25/41
Calibration 21
4.0 Calibration
The 120can be calibrated using the front panel, EDP commands, or the Revolutionconguration utility. Eachmethod consists of the following steps:
Zero calibration
Entering the test weight value Span calibration Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
Figure 4-1. Calibration (CALIBR) Menu
4.1 Front Panel CalibrationTo calibrate the indicator using the front panel, do thefollowing:
1. Place the indicator in conguration mode(display reads CONFIG) and remove all weightfrom the scale platform. If your test weights
require hooks or chains, place the hooks orchains on the scale for zero calibration.2. Press until the display reads CALIBR (see
Figure 4-1). Press to go to zero calibration(WZERO).
3. With WZERO displayed, press to calibratezero. The indicator displays *CAL*whilecalibration is in progress. When complete, theA/D count for the zero calibration is displayed.DO NOT adjust this value after WSPAN hasbeen set! Press again to save the zerocalibration value and go to the next prompt(WVAL).
4. With WVALdisplayed, place test weights on thescale and press to show the test weight value.Use the procedure shown in Figure 4-2to enterthe actual test weight, then press to save thevalue and go to span calibration (WSPAN).
5. With WSPAN displayed, press to calibratespan. The indicator displays *CAL*whilecalibration is in progress. When complete, theA/D count for the span calibration is displayed.Press again to save the span calibration valueand go to the next prompt (REZERO).
6. The rezero function is used to remove acalibration offset when hooks or chains are usedto hang the test weights. If no other apparatus was used to hang the test
weights during calibration, remove the test
weights and press to return to the CALIBRmenu. If hooks or chains were used during
calibration, remove these and the test weightsfrom the scale. With all weight removed,press to rezero the scale. This functionadjusts the zero and span calibration values.The indicator displays *CAL*while the zeroand span calibrations are adjusted. Whencomplete, the adjusted A/D count for the zerocalibration is displayed. Press to save thevalue, then press to return to the CALIBRmenu.
7. Press until the display reads CONFIG, thenpress then to exit conguration mode.
Figure 4-2. Editing Procedure for Numeric Values
WZERO
*CAL*Display and editzero calibrationA/D co unt value Display and edittest w eight value
WVAL
*CAL*Display and editspan calibrationA/D count value
WSPAN
*CAL*Press Enter to
remove offset fromzero and spancalibrations
REZERO
VERSDATET IM EPFORMTSERIALCALIBRC O NFIG FO R M AT P R O G R M
10000
When editing numeric values, press or to change the
digit selected. Press or to increment or decrement the
value of the selected digit.
P ress to save the value entered and return to the level above.
000 0 0 0
8/12/2019 120 Manual de Servicio
26/41
22 120 Installation Manual
4.2 EDP Command CalibrationTo calibrate the indicator using EDP commands, theindicator EDP port must be connected to a terminal orpersonal computer. See Section 2.2.1 on page 6forEDP port pin assignments; see Section 5.0 on page 23for more information about using EDP commands.
Once the indicator is connected to the sending device,do the following:
1. Place the indicator in conguration mode(display reads CONFIG) and remove all weightfrom the scale platform. If your test weightsrequire hooks or chains, place the hooks orchains on the scale for zero calibration.
2. Send the WZERO EDP command to calibratezero. The indicator displays *CAL*whilecalibration is in progress.NOTE: During EDP command calibration, the
*CAL*message rem ains on the disp lay. The OK
response is returned w hen calibration is comp lete.
3. Place test weights on the scale and use theWVAL command to enter the test weight valuein the following format:
WVAL= nnnnnn
4. Send the WSPAN EDP command to calibratespan. The indicator displays *CAL*whilecalibration is in progress.
5. To remove an offset value, clear all weight fromthe scale, including hooks or chains used to hangtest weights, then send the REZERO EDPcommand. The indicator displays *CAL*whilethe zero and span calibrations are adjusted.
6. Send the KUPARROW EDP command to exitconguration mode.
8/12/2019 120 Manual de Servicio
27/41
EDP Commands 23
5.0 EDP Comma nds
The 120indicator can be controlled by a personalcomputer or remote keyboard connected to theindicator EDP port. Control is provided by a set ofEDP commands that can simulate front panel key
press functions, display and change setup parameters,and perform reporting functions. The EDP portprovides the capability to print conguration data or tosave that data to an attached personal computer. Thissection describes the EDP command set andprocedures for saving and transferring data using theEDP port.
5.1 The EDP Command SetThe EDP command set includes key press commands,mode commands, reporting commands, the RS specialfunction command, and parameter setting commands.
The indicator responds to most EDP commands bysending the message OK. The OKresponse veriesthat the command was received and has beenexecuted. (Pressing ENTERafter processing a validEDP command repeats the previous command.) If thecommand is unrecognized or cannot be executed, theindicator responds with ??.
The following sections list the commands andcommand syntax used for each of these groups.
5.1.1 Key Press Commands
Key press EDP commands (see Table 5-1) simulatepressing the front panel indicat or keys. These
commands can be used in both conguration andweighing mode.
5.1.2 Mode Commands
The Mode EDP commands (Table 5-2) can be used tos w i tch th e ac t iv e mo d e o f th e in d ica to r toconguration or test mode.
5.1.3 Reporting Commands
Reporting commands (Table 5- 3) send specicinformation to the EDP port. These commands can beused in both conguration mode and normal mode.
5.1.4 The RS Command
The RS (reset conguration) command can be used torestore all conguration parameters to their defaultvalues. Before issuing this command, the indicatormust be placed in test mode (press and hold setupswitch for approximately three seconds to enter testmode).
This command is equivalent to using the DEFLTfunction on the TEST menu. See Section 7.7 onpage 35for more information about test mode andusing the TEST menu. NOTE:All load cell calibrationsettings are lost when the RS command is run.
5.1.5 Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular congurationparameter (Tables 5-4through 5-9).
Current conguration parameter settings can bedisplayed in either conguration mode or normalmode using the following syntax:
command
Mo s t p a rame te r v a lu es can b e ch an g ed inconguration mode only. Use the following commandsyntax when changing parameter values:
command= value
where valueis a number or a parameter value. Use nospaces before or after the equal (=) sign. If you type an
incorrect command or value, the display reads ??.Changes to the parameters are saved as they areentered but typically do not take effect until you exitconguration mode.
For example, to set the motion band parameter to 5,type the following:
MOTBAND= 5D
C ommand Function
K ZER O In weighing mode, press the ZER O key
KG RO SSNET In weighing mode, press the
G RO SS/NET key
K TAR E In weighing mode, press the TA RE key
K U NIT S In weighing mode, press the U NIT S k ey
K P RIN T In weighing mode, press the P RIN T key
Table 5-1. Key Press EDP Commands
C ommand Function
CO NFIG Switch to configuration mode
T EST Switch to test mode
Table 5-2. Mode EDP Commands
C ommand Function
DUM PALL List all parameter values
VERSION Write 120software version
P Write current displayed weight with units
identifier
R S Reset software
Table 5-3. EDP Reporting Commands
8/12/2019 120 Manual de Servicio
28/41
24 120 Installation Manual
C ommand Description Values
PG A Programmable gain amplifier 1, 2, 4, 8
G RADS G raduations 1100 000
ZTRK BND Zero track band O FF, 0.5D, 1D, 3D
ZRANG E Zero range 1.9% , 100%
M O TBAND M otion band 1D, 2D, 3D, 5D, 10D, 20D, 50D, O FF
O VR LO AD O verload FS+2% , FS+1D, FS+9D, FS
SM PRAT Sample rate 15HZ, 7.5HZ, 3.75HZ, 30HZ
DIGFLTR1
DIGFLTR2
DIGFLTR3
Digital filtering 1, 2, 4, 8, 16, 32, 64, 128
D FSEN S D igital filter cutout sensitivity 2O U T, 4O U T, 8O U T, 16O U T, 32O U T, 64O U T, 128O U T
D FT HR H D igital filter cutout threshold N O NE, 0.1D D, 0.2D D, 0.5D D, 1D D, 2D D, 5D D, 10D D, 20D D,
50DD , 100DD , 200DD , 250DD
Table 5-4. CONFIG EDP Commands
C ommand Description Values
PR I.DECPN T Primary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888
P R I.D SP D IV P rimary units display divisions 1D , 2D , 5D , 10D , 20D , 50D
PR I.UNITS Primary units LB, K G , O Z, TN , T, G , NO NE
SEC.D ECP NT Secondary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888
SEC.DSPDIV Secondary units display divisions 1D, 2D, 5D, 10D, 20D, 50D
SEC .UNITS Secondary units LB, K G , O Z, TN , T, G , NO NE
S EC . M U LT S econdary units multiplier 0.000009999 .99
DSPRATE Display rate 250M S, 500M S, 750M S, 1SEC , 1500M S, 2SEC , 2500M S, 3SEC ,4SEC, 6SEC, 8SEC
Table 5-5. FORMAT EDP Commands
C ommand Description Values
WZER O Zero calibration
WVAL Test weight value test_weight_value
WSPAN Span calibration
REZERO Rezero
LC .C D Set deadload coefficient value
LC .C W Set span coefficient value
Table 5-6. CALIBR EDP Commands
8/12/2019 120 Manual de Servicio
29/41
EDP Commands 25
C ommand Description Values
EDP.BAUD EDP port baud rate 1200, 2400, 4800, 9600, 19200, 38400
EDP.BITS EDP port data bits/parity 8NO NE, 7O DD, 7SPAC E, 7EVEN
EDP.T ER M EDP port termination character C R/LF, C R
EDP.EC HO EDP port echo O N, O FF
PRN.BAUD Printer port baud rate 1200, 2400, 4800, 9600
P RN.B ITS P rinter port data bits/parity 8NO NE, 7O DD, 7SPAC E, 7EVEN
P R N .T ER M P rinter port termination character C R /LF, C R
STREAM Streaming port O FF, EDP, PR N
STM DLY Stream delay NO NE, 250M S, 500M S, 1SEC , 2SEC , 4SEC , 8SEC , 15SEC
PRNDEST Print destination EDP, PRN
PRO TC T EDP port protection ENABLE, D ISABLE
Table 5-7. SERIAL EDP Commands
C ommand Description Values
PWRUPM D Power up mode G O , DELAY
REG ULAT Regulatory compliance NTEP, O IM L, C ANADA, NO NE
C ONSNUM C onsecutive number 0999 999
CO NSTUP Consecutive number start-up value 0999 999
Table 5-8. PROGRM EDP Commands
C ommand Description Values
WWPF Print characters of format string See Section 6.0 on page 27 for detailed information
WPF Print hex values of format string
Table 5-9. PFORMT EDP Commands
8/12/2019 120 Manual de Servicio
30/41
26 120 Installation Manual
5.2 Saving and Transferring DataConnecting a personal computer to the 120EDP port allows you to save indicator conguration data to the PC orto download conguration data from the PC to an indicator. The following sections describe the procedures forthese save and transfer operations.
5.2.1 Saving Indicator Data to a Personal Computer
Conguration data can be saved to a personal computer connected to the EDP port. The PC must be running acommunications program such as PROCOMMPLUS. See Section 2.2.1 on page 6for information about serialcommunications wiring and EDP port pin assignments.
When conguring the indicator, ensure that the values set for the BAUD and BITS parameters on the SERIALmenu match the baud rate, bits, and parity settings congured for the serial port on the PC. Set the PRNDESparameter to EDP.
To save all conguration data, place the indicator in conguration mode and send the DUMPALL EDP commandto the indicator. The 120responds by sending all conguration parameters to the PC as ASCII-formatted text.
5.2.2 Downloading Configuration Data from PC to Indicator
Conguration data saved on a PC or oppy disk can be downloaded from the PC to an indicator. This procedureis useful when a number of indicators with similar congurations are set up or when an indicator is replaced.
To download conguration data, connect the PC to the EDP port as described in Section 5.2.1. Place the indicatorin conguration mode and use the PC communications software to send the saved conguration data to theindicator. When transfer is complete, calibrate the indicator as described in Section 4.0 on page 21.
NOTES:
Calibration settings are included in the conguration data downloaded to the indicator. If the receivingindicator is a direct replacement for another 120and the attached scale is not changed, recalibration isnot required.
When downloading congurations that include changed serial communications settings, edit the data leto place the serial communications changes at the end of the le. Communication between the PC andindicator will be lost once the indicator receives settings for baud rate (BAUD parameter) or data bits andparity (BITS parameter) that do not match those congured for the PC.
8/12/2019 120 Manual de Servicio
31/41
Print Formatting 27
6.0 Print Formatting
The 120print format can be edited to specify theformat of the printed output when the PRINTkey ispressed or when a KPRINT EDP command isreceived.
Each print format can be customized to include up to300 characters of information, such as company nameand address, on printed tickets. You can use theindicator front panel (PFORMT menu), EDPcommands, or the Revolution conguration utility tocustomize the print format.
6.1 Print Formatting CommandsTable 6-1lists commands you can use to format theprint format. Text included in the format string mustbe enclosed in quotation marks (hex 22). Textcharacters can include any ASCII character that can be
printed by the output device.
The default 120print format is shown below:@ G < C R> @ M @ T @ N @ M < C R >
NOTE: The 300-character limit of the print formatstring includes the output eld length of the printformatting commands, not the command length. Forexample, if the indicator is congured to show a
decimal point, the @G command generates an outputeld of 13 characters: the 10-character weight value(including decimal point), one space, and a two-digitunits identier.
6.2 Customizing Print FormatsThe following sections describe procedures forcustomizing the print format using the EDP port or thefront panel (PFORMT menu).
6.2.1 Using the EDP Port
With a personal computer, terminal, or remotekeyboard attached to the 120EDP port, you can usethe EDP command set to customize the print formatstring.
To view the current setting of the print format, typeWWPF (to enter ASCII text) or WPF (to enter hexvalues) then press ENTER. The indicator responds bysending the current conguration for the print format:
Use the WWPF or WPF EDP command followed byan equals sign (=) and zero (0) to edit the print formatstring.
The following example shows the commands used todene a print format string for an Eltron LP-2742
printer.NOTE:The N, AxxxxxxN, Bx, and P1 comm ands used in
the example are all Eltron p rinter-specific commands.
WWPF=0
N
A0,0,0,3,1,2,N, B lue Hills Transfer Co. @ d @ t @ C
A8,50,0,5,1,1,N, @ G
@ M A8,120,0,5,1,1,N, @ T
A8,190,0,5,1,1,N,@ N@ M
B8,260,0,3,3,7,100,B,@ G
P1
NOTE:After entering the WWPF=0 or WPF=0 comm and,you must b egin entering the p rint format. If no data is
entered, the comm and times out, result ing in a blank
format.
C ommand Description
@ G G ross weight in displayed units
@ N Net weight in displayed units
@ T Tare weight in displayed units
@ C C onsecutive number
@ M Conditional net and tare weights.
Use the @ M command in pairs to enclose the
@ N and @ T commands. If no tare is in the
system, net and tare weights are not printed.
@ t T ime
@ d Date
@ Lnn New line (nn= number of termination characters
[CR /LF or C R ] )
@ Snn Space (nn= number of spaces)*
G ross, net, and tare weights are 9 digits in length, including
sign (10 digits with decimal point), followed by a space and
a two-digit units identifier. Total field length with units
identifier is 12 (or 13) characters.
ID and consecutive number (C N) fields are 16 characters in
length, as required.
Table 6-1. Print Format Commands
8/12/2019 120 Manual de Servicio
32/41
28 120 Installation Manual
6.2.2 Using the Front Panel
If you have no access to equipment for communication through the EDP port or are working at a site where suchequipment cannot be used, you can use the PFORMT menu (see Figure 6-1) to customize the print format. Usingthe PFORMT menu, you can edit the print format string by changing the hex values of the ASCII characters inthe format string.
To edit a print format, do the following:
1. In setup mode, use the navigation keys to go to the PFORMT menu. Press to show the EDIT submenu.2. Press again to show the print format string. Use the and keys to scroll through the format. The
number position of each character is shown in the two digits at the left of the display, hex 00BF.3. To edita character, press while the character is displayed. The rightmost digit ashes, indicating that
it can be changed. Use the and keys to increment or decrement the value, or use the key to moveto the next digit. Press to save any changes and advance to the next character in the string.
4. If done, press to return to the EDIT submenu.5. To insertone or more characters, display the character position afterwhich characters are to be inserted.
Press to return to the EDIT submenu, then press to show the INSERT parameter. Press to insertone character; repeat presses to add more characters. Each press of the key adds a character at thelocation last shown under EDIT submenu and shifts all subsequent characters to the right. Insertedcharacters are assigned hex value 00 (null).
To edit inserted characters, return to the EDIT submenu and make changes as described under step 3.6. To deleteone or more characters, display the character to be deleted. Press to return to the EDIT
submenu, then press twice to show the DELETE parameter. Press to delete one character; repeatpresses to delete more characters. Each press of the key deletes a character, starting at the location lastshown under EDIT submenu, then moving left to preceding characters. Each deletion shifts allsubsequent characters to the left.
NOTE:Some characters cannot be displayed on the 120front panel (see the ASCII character chart on page 31)and are shown as blanks. The 120can send or receive any ASCII character; the character printed depends on theparticular ASCII character set implemented for the receiving device.
Figure 6-1. PFORMT Menu, Showing Alphanumeric Character Entry Procedure
EDIT
VERSDATET IMEP F OR M TSERIALCALIBRC O NFIG FO R M AT P R O G R M
IN S ER T D ELET E
00_@40 00
0 1 _ N 4 E
Position
Character
ASCIIValue
x x_ _ 0 0
8/12/2019 120 Manual de Servicio
33/41
Appendix 29
7.0 Appendix
7.1 Error Messages
7.2 Continuous Output (Stream) FormatFigure 7-1shows the continuous output format sent to the 120EDP or printer port when the STREAM parameter(SERIAL menu) is set to either EDP or PRN.
Figure 7-1. Continuous Output Data Format
Error M essage Description Solution
_ _ _ _ _ _
(bottom LED segments lit)
D isplay overflow (negative) Negative weight value too large to be
displayed (< 99999)
_ _ _ _ _ _
(middle LED segments lit)
O verload Weight value exceeds scale capacity
_ _ _ _ _ _
(top LED segments lit)
D isplay overflow (positive) Positive weight value too large to be
displayed (> 999999)
Table 7-1. 120Error Messages
ASCII 02(decimal)
Polarity: = Positive = Negative
Weight data: 7 digits, right-justified, withdecimal point, leading zero suppression.
Error = - - - - - -
L = poundsK = kilogramT = tonG = gramsO = ounces
= none
G = GrossN = Net
Status: = validI = Invalid
M = In motion
or
ASCII 13, 10(decimal)
< / >
ASCII 32(decimal)
ASCII 32(decimal)
ASCII 47(decimal)
T ime: 8 digits, hh:mm:sshh: 0024
mm : 0059ss: 0059
Date: 2 digits, dddd : 0131
Raw count: 8 digits0000000016777216
8/12/2019 120 Manual de Servicio
34/41
30 120 Installation Manual
7.3 Front Panel Display CharactersFigure 7-2shows the 7-segment LED character set used to display alphanumeric characters on the 120frontpanel.
Figure 7-2. 120Display Characters
%
&
(
)
*
+
,
-
. .
/
0
1
2
3
4
5
6
7
8
9
:
;
?
@
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
R
S
T
U
V
W
X
Y
Z
[
\
8/12/2019 120 Manual de Servicio
35/41
Appendix 31
7.4 ASCII Character ChartUse the decimal values for ASCII characters listed inTables 7-2and 7-3when specifying print formatstrings on the 120PFORMT menu. The actualcharacter printed depends on the character mappingused by the output device.
The 120can send or receive any ASCII charactervalue (decimal 0255), but the indicator display islimited to numbers, upper-case, unaccented letters,and a few special characters.
Table 7-2. ASCII Character Chart (Part 1)
C ontrol ASC II Dec Hex ASC II Dec Hex ASC II Dec Hex ASC II Dec Hex
C trl-@ NUL 00 00 space 32 20 @ 64 40 ` 96 60
C trl-A SO H 01 01 ! 33 21 A 65 41 a 97 61
C trl-B STX 02 02 34 22 B 66 42 b 98 62
C trl-C ETX 03 03 # 35 23 C 67 43 c 99 63
C trl-D EO T 04 04 $ 36 24 D 68 44 d 100 64
C trl-E ENQ 05 05 % 37 25 E 69 45 e 101 65
C trl-F AC K 06 06 & 38 26 F 70 46 f 102 66
C trl-G BEL 07 07 39 27 G 71 47 g 103 67
C trl-H BS 08 08 ( 40 28 H 72 48 h 104 68
C trl-I HT 09 09 ) 41 29 I 73 49 i 105 69
C trl-J LF 10 0A * 42 2A J 74 4A j 106 6A
C trl-K VT 11 0B + 43 2B K 75 4B k 107 6B
C trl-L FF 12 0C , 44 2C L 76 4C l 108 6C
C trl-M C R 13 0D - 45 2D M 77 4D m 109 6D
C trl-N SO 14 0E . 46 2E N 78 4E n 110 6E
C trl-O SI 15 0F / 47 2F O 79 4F o 111 6F
C trl-P DLE 16 10 0 48 30 P 80 50 p 112 70
C trl-Q DC 1 17 11 1 49 31 Q 81 51 q 113 71
C trl-R DC 2 18 12 2 50 32 R 82 52 r 114 72
C trl-S DC 3 19 13 3 51 33 S 83 53 s 115 73
C trl-T DC 4 20 14 4 52 34 T 84 54 t 116 74
C trl-U NAK 21 15 5 53 35 U 85 55 u 117 75
C trl-V SYN 22 16 6 54 36 V 86 56 v 118 76
C trl-W ETB 23 17 7 55 37 W 87 57 w 119 77
C trl-X C AN 24 18 8 56 38 X 88 58 x 120 78
C trl-Y EM 25 19 9 57 39 Y 89 59 y 121 79
C trl-Z SUB 26 1A : 58 3A Z 90 5A z 122 7A
C trl-[ ESC 27 1B ; 59 3B [ 91 5B { 123 7B
C trl-\ FS 28 1C < 60 3C \ 92 5C | 124 7C
C trl-] G S 29 1D = 61 3D ] 93 5D } 125 7D
C trl-^ RS 30 1E > 62 3E ^ 94 5E ~ 126 7E
C trl-_ US 31 1F ? 63 3F _ 95 5F DEL 127 7F
8/12/2019 120 Manual de Servicio
36/41
8/12/2019 120 Manual de Servicio
37/41
Appendix 33
7.5 Conversion Fac tors for Secondary UnitsThe 120has the capability to mathematically convert aweight into many different types of units and instantlydisplay those results with a press of the UNITSkey.
Secondary units can be specied on the FORMATmenu using the SECNDR parameter, or by using EDPcommands.
To congure secondary units using the frontpanel menus, use the Table 7-4to nd theconversion multiplier for the MULT parameter.For example, if the primary unit is pounds andthe secondary unit is short tons, set the MULTparameter to 000050.Next, use the MULEXP parameter to set thedecimal point position. In the example above,the conversion factor for pounds to short tons isactually 0.0005 (2000 lb x 0.0005 = 1 tn). Usethe and keys to shift the decimal point toshow a value of 0.00050.Another example: If the primary unit is ouncesand the secondary unit is grams, the conversionfactor is 28.3495. To enter this value using themenus, rst enter 283495 for the MULTparameter, then use the MULEXP parameter toadjust the decimal point to 28.3495.
To congure secondary units using EDPcommands, use the Table 7-4 to nd theconversion value for the SEC.MULT andcommand. For example, if the primary unit ispounds and the secondary unit is short tons, sendthe following EDP command to set themultiplier for the secondary units.
SEC.M ULT= 0.00050
Units of weight other than those listed inTable 7-4cannot be directly specied as primaryor secondary units on the 120indicator. Forunlisted units of weight, specify NONE on theUNITS parameter.
NOTE:Ensure that the secondary decimal pointposition is set appropriately for the scale capacity inthe secondary units. If the converted value requiresmore digits than are available, the indicator displaywill overow.
For example, if the primary units are short tons,
secondary units are pounds, and the secondarydecimal point is set to 8888.88, the indicator willoverow if 5 tons or more are applied to the scale.With 5 tons applied, and a conversion factor of 2000,the secondary units display needs ve digits to the leftof the decimal point to display the 10000 lb secondaryunits value.
Primary Unit x Multiplier Secondary Unit
ounces (oz) 28.3495 grams
0.06250 pounds
0.02835 kilograms
pounds (lb) 453.592 grams
16.0000 ounces
0.45359 kilograms
0.00050 short tons (tn)
0.00045 metric tons (t)
short tons (tn) 2000.00 pounds
907.185 kilograms
0.90718 metric tons (t)
grams (g) 0.03527 ounces
0.00220 pounds
0.00100 kilograms
kilograms (kg) 35.2740 ounces
1000.00 grams
2.20462 pounds
0.00110 short tons (tn)
0.00100 metric tons (t)
metric tons (t) 2204.62 pounds
1000.00 kilograms
1.10231 short tons (tn)
Table 7-4. Conversion Factors
8/12/2019 120 Manual de Servicio
38/41
34 120 Installation Manual
7.6 Digital FilteringThe 120uses averaged digital ltering to reduce theeffect of vibration on weight readings. Adjustablethreshold and sensitivity functions allow quicksettling by suspending lter averaging, allowing theweight reading to jump to the new value. Digital
ltering parameters are set using the CONFIG menu.7.6.1 DIGFLxParameters
The rst three digital ltering parameters, DIGFL1,DIGFL2, and DIGFL3, are congurable lter stagesthat control the effect of a single A/D reading on thedisplayed weight. The value assigned to eachparameter sets the number of readings received fromthe preceding lter stage before averaging.
The overall ltering effect can be expressed by addingthe values assigned to the three lter stages:
DIGFL1 + DIGFL2 + DIGFL3
For example, if the lters are congured asDIGFL1=4, DIGFL2=8, DIGFL3=8, the overall l te r ing e f fec t i s 20 (4 + 8 + 8 ) . With th isconguration, each A/D reading has a 1-in-20 effecton the displayed weight value. Setting the lters to 1effectively disables digital ltering.
7.6.2 DFSENS a nd DFTHRH Pa rameters
The three digital lters can be used by themselves toeliminate vibration effects, but heavy ltering alsoincreases settling time. The DFSENS (digital ltersensitivity) and DFTHRH (digital lter threshold)parameters can be used to temporarily override lter
averaging and improve settling time: DFSENS species the number of consecutivescale readings that must fall outside the lterthreshold (DFTHRH) before digital lteringis suspended.
DFTHRH sets a threshold value, in displaydivisions. When a specied number ofconsecutive scale readings (DFSENS) falloutside of this threshold, digital ltering issuspended. Set DFTHRH to NONE to turn offthe lter override.
7.6.3 Setting the Digital Filter Parameters
Fine-tuning the digital lter parameters greatlyimproves indicator performance in heavy-vibrationenvironments. Use the following procedure todetermine vibration effects on the scale and optimizethe digital ltering conguration.
1. In setup mode, set all three digital lters(DIGFL1, DIGFL2, DIGFL3) to 1. SetDFTHRH to NONE. Return indicator tonormal mode.
2. Remove all weight from the scale, then watchthe indicator display to determine themagnitude of vibration effects on the scale.Record the weight below which all but a fewreadings fall. This value is used to calculatethe DFTHRH parameter value in Step 4.For example, if a heavy-capacity scaleproduces vibration-related readings of up to50 lb, with occasional spikes to 75 lb, record
50 lb as the threshold weight value.3. Place the indicator in setup mode and set thedigital lters (DIGFLx) to eliminate thevibration effects on the scale . (LeaveDFTHRH set to NONE.) Recongure asnecessary to nd the lowest effective valuesfor the DIGFLxparameters.
4. With optimum values assigned to the DIGFLxparameters, calculate the DFTHRH parametervalue by converting the weight value recordedin Step 2 to display divisions:threshold_weight_value / DSPDIV
In the example in Step 2, with a thresholdweight value of 50 lb and a display divisionvalue of 5lb: 50 / 5lb = 10DD. DFTHRHshould be set to 10DD for this example.
5. Finally, set the DFSENS parameter highenough to ignore transient peaks. Longertransients (typically caused by lower vibrationfrequencies) will cause more consecutiveout-of-band readings, so DFSENS should beset higher to counter low frequency transients.Recongure as necessary to nd the lowesteffective value for the DFSENS parameter.
8/12/2019 120 Manual de Servicio
39/41
Appendix 35
7.7 Test ModeIn addition to normal and setup modes, test modeprovides a number of diagnostic functions for the 120,including:
Display raw A/D count Display digital lter raw counts Reset conguration parameters to default values Transmit test character (U) from serial port Display characters received by serial port
To enter test mode, press and hold the setup switch.After about three seconds, the test mode displayautomatically shifts to the rst test menu function,A/DTST.
Figure 7-3shows the Test Menu structure. Note that,because the Test Menu functions are all on a singlemenu level, the GROSS/NET( ) key has no function.Press the ZERO( ) key to exit test mode.
Table 7-5 on page 35summarizes the test menufunctions.
Figure 7-3. Test Menu
TEST Menu
Function Description
A/DTST D isplay A /D test
Press and hold Enter key to display raw count from A/D converter.
FILTR1
FILTR2
FILTR3
Display filtered raw count for digital filters 13
DEFLT Default parameters
Press setup switch and Enter key at the same time to reset configuration and calibration parameters to factory
default values. Load cells must be recalibrated before using the indicator (see Section 4.0 on page 21).
XM T U Transmit U
Press and hold Enter key to send ASC II U characters (decimal 85) from the serial port.
ECH O R Echo received characters
When R eady is displayed, press the Enter key to view characters received at serial
port. The first two digits show the order of the character received (see diagram(.
Table 7-5. Test Menu Functions
A/DTST FILTR 1 FILTR 2 FILTR 3
DEFLT XM T U ECHO R
0 1 _ N 4 E
OrderReceived
Character
ASCIIValue
8/12/2019 120 Manual de Servicio
40/41
36 120 Installation Manual
7.8 Spec ifica tions
Power
Line Voltages 115 or 230 VAC
Frequency 50 or 60 Hz
Power Consumption 15 mA @ 115 VAC (2 W)
8 mA @ 230 VAC (2 W)
M ain C ircuit
Input Voltage 8.012.0 VD C
M aximum Current 250 mA
Analog Specifi ca tions
Full Scale Input Signal Up to 22.5 mV
Excitation Voltage 5 0.3 VDC,
4 x 350Wor 8 x 700Wload cells
Sense Amplifier Differential amplifier with
4- and 6-wire sensing
Analog Signal Input Range (zero load)
PG A = 1: 0.5 mV/V to +0.5 mV/V
PG A = 2: 0.2 mV/V to +0.2 mV/V
PG A = 3: 0.1 mV/V to +0.1 mV/V
PG A = 4: 0.05 mV/V to +0.05 mV/V
Analog Signal Input Range (full load)
P G A = 1: 4.5 mV/V
P G A = 2: 2.2 mV/V
P G A = 3: 1.1 mV/V
P G A = 4: 0.6 mV/V
Analog Signal
Sensitivity 0.15 mV/graduation minimum,1.0 mV/grad recommended
Input Impedance 200 M W, typical
Noise (ref to input) 0.3 mV p-p with digital filters at 4-4-4
Internal Resolution 1 677 000 counts
Display Resolution 999 999 dd
M easurement Rate Up to 30 measurements/sec
Input Sensitivity 0.6 nV per internal count
System Linearity Within 0.01% of full scale
Zero S tability 150 nV/C , maximum
S pan S tability 3.5 ppm/C , maximum
C alibration M ethod Software, constants stored in
EEPROM
C ommon M ode
Voltage +1.5 to +3.5 V, referred to zero voltage
C ommon M ode
R ejection 130 dB minimum @ 50 or 60 H z
Normal M ode
R ejection 90 dB minimum @ 50 or 60 H z
IRFI Protection Signal, excitation, and sense lines
protected by capacitor bypass
Digital Specifi ca tions
M icrocomputer Intel M CS-52 with 16K EEPROM @
18.432 M Hz
Digital Filters 3 filters, software selectable
Serial Communications
ED P P ort Full duplex R S -232, 120038400 bps
Printer Port O utput-only RS-232 or active 20 mA
current loop, 12009600 bps
Both Ports 7 or 8 data bits; even, odd, space, or no
parity
Operator Interface
D isplay 6-digit LED display. 7-segment, 0.8 in
(20 mm) digits
LED annunciators G ross, net, center of zero, standstill,
lb/primary units, kg/secondary units,
oz, g
K eyboard 5-key flat membrane panel
Environmental
O perating Temperature10 to +40C (legal);
10 to + 50C (industrial)
Storage Temperature 25 to +70C
Humidity 095% relative humidity
Altitude 2000 m (6500 ft) maximum
Enclosure
Enclosure Dimensions 7.3 in x 3.7 in x 3.9 in
18.6 cm x 9.5 cm x 10 cm
Weight 0.9 lb (0.4 K g)
R ating/M aterial ABS plastic
Certific ations a nd Approvals
NTEP
C oC Number 03-059
A ccuracy C lass III/IIIL
nmax
: 6 000
Measurement Canada
Approval
Accuracy Class
nmax
: 10 000
OIML
R76-2 Test Certificate
Accuracy C lass III nmax
: 3 000
NAT
IONA
L CONFERENC
E
ON
WE
IGH
TS AND MEA
SUR
ES
8/12/2019 120 Manual de Servicio
41/41
120 Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by aDistributor or Original Equipment Manufacturer (OEM) will operate per written specications as conrmed bythe Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects inmaterials and workmanship for one year.
RLWS warrants that the equipment sold hereunder will conform to the current written specications authorizedby RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipmentfails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within thewarranty period subject to the following conditions:
Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with adetailed explanation of the alleged deciencies.
Individual electronic components returned to RLWS for warranty purposes must be packaged toprevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in apublication, Protecting Your Components From Static Damage in Shipment, available from RLWSEquipment Return Department.
Examination of such equipment by RLWS conrms that the nonconformity actually exists, and wasnot caused by accident, misuse, neglect, alteration, improper installation, improper repair or
improper testing; RLWS shall be the sole judge of all alleged non-conformities. Such equipment has not been modied, altered, or changed by any person other than RLWS or its
duly authorized repair agents. RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible
for shipping charges both ways. In no event will RLWS be responsible for travel time or on-location repairs, including assembly or
disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES , EXPRESSED OR IMPLIED , INCLUDINGWITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULARPURPOSE . NEITHER RLWS NOR DISTRIBUTOR WILL , IN ANY EVENT , BE LIABLE FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES .
RLWS AND BUYER AGREE THAT RLWSS SOLE AND EXCLUSIVE LIABILITY HEREUNDER ISLIMITED TO REPAIR OR REPLACEMENT OF SUCH GOODS . INACCEPTING THIS WARRANTY , THE
BUYER WAIVES ANY AND ALL OTHER CLAIMS TO WARRANTY .
SHOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE SELLERFOR WARRANTY CLAIMS .
NO TERMS , CONDITIONS , UNDERSTANDING , OR AGREEMENTS PURPORTING TO MODIFY THETERMS OF THIS WARRANTY SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN WRITING AND
SIGNED BY ACORPORATE OFFICER OF RLWS AND THE BUYER .